Programming - Sirona Orthophos 3 Gebrauchsanweisung

Panorama-röntgengeräte
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7 Programming

7
Programming
P I
66
I0
3.
04
28
kV/mA values have been assigned to the four program
buttons at the factory.
For free programming the buttons can be programmed
with different values.
See chapter 'Program Values'.
1.
Programming Procedure
1.
Push buttons – + to select program number P1... to
2.
be changed.
2.
Push – + buttons to set desired kV value on the dig-
ital display.
3.
Push memory button. The LED over the memory
4.
button blinks.
4.
Push the patient symbol button to be repro-
grammed. The LED above this button lights up.
The LED above the memory button is switched off.
The new values are now stored.
Programming is complete.
Please enter the new value in the 'Freely programmed
values' table, see page 30.
Adjustment of the Exposure
The adjustment of the exposure is set to 03 at the fac-
tory.
If the degree of exposure is to be changed, use the sup-
plied screwdriver as shown in the illustration.
During adjustment the corresponding switch position is
briefly displayed
(example 04).
Step switch set to– =
Step switch set to+ =
i
NOTE
+
Changing the film density automatically alters the pro-
grammed kV/mA values. Refer to 'Program Values', see
page 30.
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS 3
lower dose,
lighter exposure
higher dose,
darker exposure
60 82 262 D 3285
D 3285.201.04.06.02