3.6

Thermodisinfecting the suction hoses

Grease
59 06 511 D 3386
D 3386.103.01.06.02
The unit is equipped as standard with suction hoses that
cannot be thermodisinfected.
However, it may have suction hoses that can be
thermodisinfected as special accessories. These are
Grease
then marked in turquoise color (10).
ATTENTION
Only hoses marked with a turquoise ring (10) can be
thermodisinfected!
First switch the treatment center OFF.
10
Pull off the suction hoses and the connection hose from
the assistant element and the water unit.
Remove the top parts (5) of the filter housings and the
amalgam residue collector (6).
Pull off the handpiece (7) from the suction hose.
Grease
For the saliva ejector hose, insert the intermediate piece
(8) in the hose holder (9).
Grease the O-rings.
ATTENTION
Use only lubricants which have been approved by
Sirona!
Fit the hose holders onto the rails in the
thermodisinfector (by Miele) as shown.
Connect the suction hoses to the hose holders and place
them in the wire basket (by Miele) for thermodisinfection.
Thermodisinfect at max. 93 °C.
Before reconnecting the hoses to the assistant element,
lubricate the O-rings at the connection pieces .
The hoses must engage firmly.
Hose holder for
Miele thermodisinfector
Order No. 89 18 757