6 Disinfection system
6.2 Sanitation
San.
Sanier – Phase
Stop
(Sanitation phase)
1
2
3
4
20
1
2
2
3
4
5
6
Sanitation phase 2
First the program checks whether maximum water flow
has been set for the instruments.
If there is no water flow, an acoustic signal sounds. The
symbol of the corresponding receptacle flashes and the
display shows the symbol:
(Example receptacle 3 on the dentist element).
In this case open the water regulator of the
corresponding hose on the instrument coupling. Leave
the relevant hose plugged into the adapter in this case.
As soon as the system detects that water is flowing,
sanitation is continued automatically.
i
NOTE
If you press the flashing key after the acoustic signal
sounds without opening the water regulator, the
sanitation program continues. The corresponding
water path is not sanitized.
Sanitation phase 3
The water is pumped out from the water tank of the water
unit via the tumbler filler.
Sanitation phase 4
The container is now automatically filled with
DENTOSEPT P.
This is then used for rinsing the instrument hoses, the
SPRAYVIT hoses with valve bodies and the tumbler
filler. After the individual rinsing process the
corresponding receptacle symbol appears dark.