Sirona Dental Systems GmbH
6.2

Sanitation

Special measures for additional devices
Start
San.
12:30
59 06 511 D 3386
D 3386.103.01.06.02
0m
42s
C
Desinf.
Special devices that are connected to the chair via the
media block are treated with disinfected water with a
higher hydrogen peroxide concentration (1.4%) during
sanitation.
ATTENTION
If the connected devices are not suitable for this higher
sanitation concentration, Sirona recommends to
disconnect them from the treatment center during
sanitation and to perform a separate sanitation in
accordance with the manufacturer's specifications.
Preparation of sanitation
Perform sanitation every 4 weeks, or if the
germination index significantly exceeds 100 germs
per ml (see section "Microbiological water test").
Only in this way can you surely and effectively
combat the formation of biofilm in the water paths.
• It takes approx. 10 minutes to prepare sanitation.
• If Desinf. flashes on the display of the dentist element,
DENTOSEPT P must be filled in first (see Operating
Instructions).
ATTENTION
In Japan the disinfectant OXYDOL available there must
be used instead of DENTOSEPT P.
OXYDOL must be mixed 1:1 with distilled water before it
is filled into the disinfection system.
The flashing Desinf message then disappears.
DENTOSEPT P
1 carton = 6x1 liter, Order No. 33 18 156
• Maximum water flow must be set on all drill drives as
well as on the SIROSONIC L. Check this before every
sanitation.
• All instruments and suction hoses must then be placed
in their receptacles.
6 Disinfection system
6.2 Sanitation
17