1 Connection box

1
Connection box
6
Should you notice changes of the media flows, please
check the filters for water and air (1) for permeability.
Replace them if necessary.
Replace the filters (1) for water and air.
To do so, first turn the main switch OFF. Pry off the cover
(2) with a screwdriver.
Remove the screw caps (3) and replace the filters.
OFF
Filter
Order No. 14 43 436
The main fuse (4) is located below the air filter.
Unit main fuse for 230V AC:
T 6.3A H, 250V AC, Order No. 10 77 452
Unit main fuse for 100V AC / 115V AC:
T 10A H, 250V AC, Order No. 10 77 460
Sirona Dental Systems GmbH
59 06 511 D 3386
D 3386.103.01.06.02