After each patient
Clean the cuspidor.
Use only cleaning agents which have been approved by
Amalgam
Sirona!
Do not use any domestic cleaning and care agents.
Foaming domestic cleaning agents lead to foam and
6
water being drawn into the dry aspiration system,
resulting in damage to the suction machine.
Daily before the mid-day break and
after the practice has closed
Remove the gold trap (5) and clean it.
Store amalgam residues under water in a separate
closed container. When the amalgam rotor is replaced
(see page 14), put the residues into the rotor for
disposal.
Grease
The cuspidor must be filled with at least 1 liter of water
for rinsing after the cleaning process.
Amalgam residues in the gold trap must not be disposed
of in the cuspidor or the sink!
Weekly
Pull out the tumbler outlet (6) or the hydrocolloid
adapter, if any, and clean it.
Before refitting the tumbler outlet or the hydrocolloid
adapter, lightly lubricate the O-rings.
Use only lubricants which have been approved by
Sirona!
ATTENTION
ATTENTION
ATTENTION
ATTENTION
2 Cuspidor
7