Herunterladen Inhalt Inhalt Diese Seite drucken

Operating Mode; Temperature Control Mode - IKA C-MAG HS 7 control Betriebsanleitung

Vorschau ausblenden Andere Handbücher für C-MAG HS 7 control:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

/// Operating mode

Operating the device in mode A, B or D.
Mode A
All settings will be stored if the device is switched off or disconnected from
the power supply. The stirring and heating functions will be set to OFF when
the device is powered on. The safety circuit can be set or modified. When
the power supply switch is switched on, A is displayed.
Mode B
All settings will be stored if the device is switched off or disconnected from
the power supply. The stirring and heating functions will be set to ON or
OFF when the device is powered on, depending on the previous status
of the device. The safety circuit can be set or modified. When the power
supply switch is switched on, B is displayed.
Mode D
Switch the device on
The safe temperature has to be confirmed by pushing one of the
rotating knobs (C or D).
During operation
In Mode D the device behaves the same as in Mode A with the
exception that:
› The temperature / speed setting must be confirmed by pressing the
temperature / speed knob. To change the temperature / speed setting,
turn the temperature / speed knob until the desired value appears.
› The new value flashes for 5 seconds on the display. Confirm the new
temperature / speed setting by pressing the temperature / speed knob,
otherwise the temperature / speed setting value will skip back to its
previous value.
86
Menu
T
Set

/// Temperature control mode

When using an external PT 1000 temperature sensor, you can choose
between two types of control:
PID mode
Good control results, minimized overshooting, slow rise in temperature.
T
Set
2P-mode (two-point controller)
Maximum heating rate, increased overshooting.
T
Set
t
T
Set
t
t
Menu
87

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis