/// Accessories

› Safe operation is only guaranteed with the accessories
described in the "Accessories" chapter.
› Always disconnect the plug before attaching accessories.
› Observe the operating instructions of the accessories.
› Ensure that the external temperature sensor is inserted into
the medium to a depth of at least 20 mm when connected.
› Accessories must be securely attached to the device and
cannot come off by themselves. The centre of gravity of the
assembly must lie within the surface on which it is set up.

/// Power supply / Switching off

› The voltage stated on the type plate must correspond to
the mains voltage.
› The outlet for the mains plug must be easily accessible.
› The device can only be disconnected from the mains outlet
by pulling out the power cord or the connector plug.
Warning! (
)
› The device will automatically restart in mode B following
any interruption to the power supply.
› After continuously running for 100 hours the device will limit
the maximum heating power by 10% for self-protection.
70
Safety instructions

/// About protecting your device

› The stirrer may only be opened by qualified and IKA approved experts.
› Do not cover the device, even partially e.g. with
metallic plates or film. This may result in overheating.
› Protect the device and accessories from bumps and impacts.
› Keep base plate clean.
› Observe the minimum distances:
between devices min. 100 mm,
between device and wall min. 100 mm,
above the device min. 800 mm.
> 100 mm
> 100 mm
Safety instructions
71