4.2.2 Use of lithium-ion rechargeable batteries

The integrated lithium-ion battery is subject to the requirements of the
Dangerous Goods Directive. Special requirements for packaging and
labelling must be observed when shipping.
A dangerous goods expert must be consulted when preparing the package.
Do not ship the CPH7650 if the rechargeable battery is damaged or
defective.
Observe the different dangerous goods requirements relative to the
respective modes of transport and any other national regulations.
The rechargeable battery lifetime is up to 8 hours for continuous operation (without
backlighting).
The rechargeable battery is permanently installed in the model CPH7650 pressure
calibrator.
The battery can be charged only using the power supply unit included in the scope of
delivery.
In case the battery in the process calibrator stops working, contact WIKA.
The instrument must not be opened under any circumstances.
For contact details see chapter 1 "General information" or the back page of
the operating instructions.

4.2.3 Charging the lithium-ion rechargeable battery

DANGER!
Physical injuries, damage to property and the environment due to an
improper charging of the pressure calibrator!
Incorrect charging of the pressure calibrator can immediately lead to
dangerous situations and serious injuries, since the built-in lithium-ion
rechargeable battery can explode, heat up or ignite.
Only use the power supply unit from WIKA supplied with the instrument!
Never leave the pressure calibrator near fire, nor in direct sunlight. If the
lithium-ion batteries become hot, the built-in safety device is activated and
overcharging prevented. Heating the lithium-ion batteries can damage the
safety device and can thus lead them to heat up further, to cease to work
or to ignite.
Do not continue to charge the pressure calibrator if it does not fully
recharge within the specified time (approx. 8 hours).
The cover of the calibrator's case must be kept open during the charging
process!
WIKA operating instructions, model CPH7650