5.2 Setting the heating bath temperature

WARNING
CAUTION
CAUTION
1.02
Using the Laborota 4000 / 4001 efficient; 4010 / 4011 digital
1. Switch on the power switch of the instrument base (3).
The instrument base power switch will be lit green (3).
2. Set the Rotation knob (4) to the desired speed.
The flask will start rotating. The Laborota 4010 / 4011 digital shows the
selected speed on its display (1).
Heating bath knob is set to left stop.
Heating bath is filled with heating medium.
Rotary evaporator is operational and the power switch is on.
Surface temperatures in excess of 60°C during heating bath operation!
Danger of burns.
Do not touch the inside and upper edge of the heating bath, the
evaporating flask and the heating bath medium.
Wear protective gloves when changing the evaporating flask.
Overheated heating bath!
Damage and changes to the appearance of the heating bath.
Never use the heating bath without liquid.
Damage to the guard hood!
Set a maximum setpoint of 120 °C for operation with a guard hood.
The temperature difference between the heating bath and the vapour should
be at least 20 K to ensure an adequate distillation rate.
As a rule of thumb, doubling the temperature difference will double the
distillation rate.
Use a suitable oil as a heating bath medium for setpoint temperatures in
excess of 100°C (observe safety data sheets, see chapter 4.4.2, page 83).
The message OIL will appear on the display of the heating bath (4) for one
second when setting temperatures higher than 100°C.