Herunterladen Inhalt Inhalt Diese Seite drucken

Adding Distillation Material; Manual Aeration; Installing/Connecting Accessories - Heidolph Laborota 4000 efficient Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Setup and initial use

4.6 Adding distillation material

4.7 Manual aeration

4.7.1
Guard hood
102
Additional distillation material can be added during distillation and while under
vacuum via the inlet hose and pipe.
Fig. 4-23:
Inlet hose
1. Insert the inlet hose (1) in the additional material.
2. Turn the cock plug (2) parallel to the inlet hose (1).
The additional distillation material will be drawn into the evaporating flask.
The system can be aerated via the cock plug (Fig. 4-23 (2)) of the inlet pipe:
Turn the cock plug (Fig. 4-23 (2)) of the inlet pipe slowly upward.
The system will be aerated.

Installing/connecting accessories

Fig. 4-24:
Installing the guard hood
1. Attach the guard hood holder (2) to the vertical stop (1) with the two
included Allen-head screws (3).
2. Release the clamping screw (4) and align the guard hood (5) to the
heating bath by moving it horizontally.
3. Retighten the clamping screw.
The guard hood is now installed. It can be opened using the grip.
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control
1.02

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis