Manual lift
Motorised lift

4.5 Installing glassware

86
Laborota 4000 efficient, 4010 digital, 4002 control (manual lift)
Fig. 4-3:
Manual lift
1. Press the manual lift (1) downward and move left (lower) or right (higher)
while pressed.
2. Release the manual lift (1) at the desired position.
The lift will be locked at the new position when released.
Laborota 4001 efficient, 4011 digital, 4003 control (motorised lift)
Instrument base is installed.
(for Laborota 4003 control) None of the Set modes are active
(see chapter 6, page 109).
Press the appropriate arrow button (up/down) (see Fig. 6-1, page 109 (9))
and release at the desired height.
The drive will remain at the desired height.
In the Laborota 4003 control, the motorised lift will not work as long as the
instrument is in one of the Set modes.
The following section describes the installation of the glassware. The rotary
evaporator can be equipped with five different sets of glassware.
Instrument base is set up and installed.
Water connection is available (not for glassware G5).
Lift is in highest position (see chapter 4.4.5, page 85).