6.2 Bench alignment of the prosthesis

INFORMATION
The prosthetic components and prosthesis shown in the illustrations are used as examples to
illustrate the general process. The instructions for use of the prosthetic components selected for
the patient contain detailed information and have to be used for fabrication of the prosthesis.
► CAUTION! To enable the patient to stand safely, align the prosthesis using the alignment
images for the sagittal plane (see Fig. 
3
 prosthetic knee,
1
 alignment line,
ing subsections.
INFORMATION: Using an alignment apparatus 
alignment. The retaining bits 
12
(see Fig. 
on page 11).

6.2.1 Activating and deactivating lock

► NOTICE! To avoid damage to the prosthesis during deactivation and activation of the lock, put
the prosthetic joint in flexion (≥90°).
► If the prosthetic socket 
13
figure 
on page 11.
► If the prosthetic socket 
on page 12 (factory default setting).

6.2.2 Secure connection cord

► Secure the connection cord, as shown in figure 
and to the spring travel limiter.

6.2.3 Positioning the prosthetic foot

► Position the prosthetic foot as shown in Fig. 
► NOTICE! To prevent rapid wear, adhere to the tolerance range for exterior rotation and do not
adapt to the exterior rotation of the sound foot.

6.2.4 Mounting the distal adapters

CAUTION
Improper assembly of the screw connections
Fall due to breakage or loosening of the screw connections
Clean the threads before every installation.
Apply the specified installation torque values.
Observe the instructions for securing the screw connections and the use of the correct
length.
► Mount the distal adapters of the prosthetic knee joint on the prosthetic foot (see Fig. 
page 13).

6.2.5 Mounting the prosthetic knee joint

CAUTION
Incorrect mounting on the tube
Falling due to breakage of load-bearing components
Do not use the insertion range of the prosthetic knee joint for length compensation.
► CAUTION! Clean the contact surfaces of the prosthetic knee joint and tube using a degreas­
ing cleaning agent compatible with the material.
4
 prosthetic socket) and for the frontal plane (see Fig. 
2
3
 prosthetic foot,
3
are used to clamp the prosthetic knee joint in the knee bracket 
1
is used without a lock slide, deactivate the lock according to
1
is used with a lock slide, activate the lock according to figure 
10
1
on page 10 –
4
 prosthetic knee,
1
with lasers 
15
on page 12, to the prosthetic knee joint
16
on page 13.
2
 alignment line,
prosthetic foot,
11
 prosthetic socket) and the follow­
2
is recommended for bench
on page 10 –
4
14
17
on