Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radio, amateur radio, AM
and FM radio broadcast and TV broadcast, cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the MicroRPM is used
exceeds the applicable RF compliance level above, the MicroRPM should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orientating or relocating
the MicroRPM.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3 V/m
3 Vrms
3 V
150kHz to
80MHz
3 V/m
3 V/m
80MHz to
2.5GHz
Recommended separation
distance
d = 1.2 x √P
d = 1.2 x √P
800MHz
d = 2.3 x √P
2.5GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from a fixed RF
transmitters as determined by
an electromagnetic site
survey, a should be less than
the compliance level in each
frequency range. b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
50
80MHz to
800MHz to