Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The MicroRPM is intended for use in the electromagnetic environment
specified below. The customer or the user of the MicroRPM should assure
that it is used in such an environment.
Immunity
Test
+/- 1 kV line(s)
to line (s)
+/- 2 kV line(s)
to earth
< 5% UT
(>95% dip in
UT)
for 0.5% cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in
UT)
for 5s
3 A/m
IEC
Compliance
60601
Level
test level
Not Applicable
Not Applicable
3 A/m
Electromagnetic
Environment -Guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the micro RPM,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
49
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Mains power quality
should be that of a
typical commercial
or hospital
environment. If the
user of the micro
RPM requires
continued operation
during power main
interruptions, it is
recommended that
the micro RPM be
powered from an
uninterruptible
power supply or a
battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.