CareFusion MicroRPM Benutzerhandbuch Seite 25

Inhaltsverzeichnis

Werbung

Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
+/- 1 kV
line(s) to line
(s)
+/- 2 kV
line(s) to earth
< 5% UT
(>95% dip in
UT)
for 0.5% cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
(30% dip in
UT)
for 25 cycles
< 5% UT
(>95% dip in
UT)
for 5s
3 A/m
Not Applicable
Not Applicable
3 A/m
23
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the micro
RPM requires continued
operation during power
main interruptions, it is
recommended that the
micro RPM be powered
from an uninterruptible
power supply or a
battery.
Power frequency
magnetic fields should
be at levels
characteristics of a
typical location in a
typical commercial or
hospital environment.