i) Capacity Measuring

Make sure that all circuit parts, circuits and components and other objects of measurement are
disconnected from the voltage and discharged.
Always observe polarity with electrolyte capacitors.
- Turn the DMM on and select measuring function "CAP".
- Press "MODE" button 3x to switch measurement functions. The
display shows the unit "nF" for capacity measurement. Pressing this
button again takes you to the next measuring function, etc.
- Plug the red measuring line into the Ω measuring jack (7) and the
black measuring line into the COM measuring jack (8).
Due to the sensitive measuring input, the display may show a
value if the measuring lines are "open". By pressing the button
"REL", the display is set to "0". The Auto range function remains
active.
- Connect the two measuring prods (red = plus/black = minus) with the
object to be measured (capacitator). After a short time the display
shows the capacity. Wait until the displayed value has stabilised. This
may take a few seconds for capacities of >40 µF.
- If "OL" (overload) appears on the display, you have exceeded the
measurement range.
- Remove the measuring lines from the object to be measured after
completion of the measurement and switch off the DMM.