b) Measure sets

Further settings can be made in the individual measuring functions and the values to be displayed can be enabled
or disabled there.
To get to the setting menu, push the button "Enter" in a measuring function. The menu "MEASURE SET" is opened.
Select by pushing the button "Enter" again. You can use the arrows to change the function/set value. To terminate,
push the button "Esc".
The following functions can be selected and set:
Emission
The emission degree can be set individually in this menu from 0.10 to 1.00 or selected with pre-set material values.
Possible set values are:
ε
= x.xx set value 0.10 - 1.00
Cement (0.94)
Glass (0.92)
Human Skin (0.98)
Alarm High
Here, the upper alarm level (High) of the temperature can be set. You can enable or disable the function via the
selection field.
Possible set values are:
Time value = xx.x set value -50 to +1650 °C
Disable = Function is disabled
Alarm Low
Here, the lower alarm level (Low) of the temperature can be set. You can enable or disable the function via the
selection field.
Possible set values are:
Time value = xx.x set value -50 to +1650 °C
Disable = Function is disabled
Laser
Here, the target laser during IR measuring can be switched on or off. Pushing the button "Enter" actuates the slider
in the display.
I = enabled, o = disabled
Ice/Water (0.96)
Plastic (0.90)
Wood (0.87)
Enable = Function is enabled
Enable = Function is enabled