Herunterladen Inhalt Inhalt Diese Seite drucken

Preparing The Measurement - Benning MM 1-3 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für MM 1-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
Measuring
Resolution
range
2 kHz
1 Hz
20 kHz
10 Hz
200 kHz
100 Hz
2 MHz
1 kHz
20 MHz
10 kHz
7.9 Temperature ranges °C (BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
Measuring range
-20 °C ~ 0 °C
1 °C ~ 100 °C
101 °C ~ 500 °C
501 °C ~ 800 °C
7.10 Temperature ranges °F (BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
Measuring range
-4 °F ~ 32 °F
33 °F ~ 212 °F
213 °F ~ 932 °F
933 °F ~ 1472 °F
8.
Measuring by means of the BENNING MM 1‑1/ 1‑2/ 1‑3

8.1 Preparing the measurement

Operate and store the BENNING MM 1-1/ 1-2/ 1-3 at the specified storage and
operating temperatures only!
Do not permanently expose the device to sunlight.
-
Check stated nominal voltage and nominal current on the safety measuring
lines. Nominal voltage and current of the enclosed safety measuring lines
comply with the respective values of the BENNING MM 1-1/ 1-2/ 1-3.
-
Check insulation of the safety measuring lines. If the insulation is damaged,
the safety measuring lines must be replaced immediately.
-
Check the safety measuring lines for continuity. If the conductor in the
safety measuring line is interrupted, replace the safety measuring lines
immediately.
-
Before selecting another function by means of the rotary switch 8,
disconnect the safety measuring lines from the measuring point.
-
Strong sources of interference in the vicinity of the BENNING MM 1-1/ 1-2/ 1-3
might involve unstable readings and measuring errors.
8.2 Voltage and current measurement
Do not exceed the maximum permitted voltage with respect to
earth potential! Electrical danger!
The highest voltage which may be applied to the
-
COM jack J
-
jack for V, Ω,
-
jack for the 10 A range K (BENNING MM 1-2/ 1-3)
of the BENNING MM 1-1/ 1-2/ 1-3 against ground is 1000 V.
Electrical danger!
Maximum circuit voltage for a measured current of 500 V! In
case of fuse triggering at values above 500 V, damages of the
device might be involved. A damaged device might represent
an electrical hazard!
11/ 2007
Measuring accuracy for 5 V
± (0.01 % of the measured value + 1 digits)
± (0.01 % of the measured value + 1 digits)
± (0.01 % of the measured value + 1 digits)
± (0.01 % of the measured value + 1 digits)
± (0.01 % of the measured value + 1 digits)
Resolution
Resolution
, Hz 9
BENNING MM 1‑1/ 1‑2/ 1‑3
max.
eff
1 °C
1 °C
1 °C
1 °C
1 °F
1 °F
1 °F
1 °F
Sensitivity
> 1.5 < 5 V
> 1.5 < 5 V
> 1.5 < 5 V
> 2 < 5 V
> 2 < 5 V
eff
Measuring accuracy
± (2 % + 4 °C)
± (1 % + 3 °C)
± (2 % + 3 °C)
± (3 % + 2 °C)
eff
Measuring accuracy
± (2 % + 8 °F)
± (1 % + 6 °F)
± (2 % + 6 °F)
± (3 % + 4 °F)
eff
eff
eff
eff
eff
17

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mm 1-1Mm 1-2

Inhaltsverzeichnis