6.3. To Set the Screen Saver

The screen saver will run automatically if no operation is taken for any key within the set
time. Press any button to resume.
1. Press Utility and choose Disp Setup, press F3 to select Scrn Svr.
2. Press F3 to switch between On/Off.
3. If On is selected, you can set the screen saver time. Turn the knob to change the
value, press
number keys to input the desired time in minutes, press
screen saver time range is 1~999 minutes.

6.4. To Set the Counter

The counter can measure the signal with frequency form 100mHz to 200MHz.
The operation steps:
Utility
1. Press
and choose Counter.
2. Connect the signal to the connector [20MHz In/Counter] on the rear panel.
F3
3. Press
to choose Set, enter measurement setting menu.
To set the coupling mode: Press F1 to switch Coupling as AC/DC.
To set the sensitivity: Press F2 to switch Sens as Low/Middle/High.
For low amplitude signal, the "Middle" or "High" sensitivity should be used.
For low frequency signal with high amplitude and slower rising edge, low sensitivity is
a better choice.
To set the high frequency restrain on/off:
Press F3 to switch HFR as ON/OFF.
High frequency restrain is used for filtering the high frequency signal in measuring the
low frequency signal, and improve the measure accuracy.
To measure low frequency signal lower than 1kHz, you should put on the high
frequency restrain to filter the high frequency noise disturb.
To measure high frequency signal higher than 1kHz, you should put off the high
frequency restrain.
To set the trig level: Press
value, press
number keys to input the desired value and choose the unit. The range of trig levle is
0~2.5V.
Press F5 to choose Back to previous menu, view the result.
After all the set was done, the counter will measure the signal according to the current set.
If the reading is unsteady, please repeat the above step until it is steady.
4. Press F1 to switch to view the measure result of frequency or period.
Press F2 to switch to view the measure result of positive width or duty.
/
direction key to move the cursor left and right; or press the
F4 to choose TrigLev. Turn the knob to change the
/
direction key to move the cursor left and right; or press the
-94-
F4 to select the unit. The