7. To Set Output Parameter

7.1. To Set the Output Load

For either of CH1 Output and CH2 Output on the Front panel, the Generator has a built-in
50Ωseries impendence. If the actual load does not match the set one, the displayed
amplitude and offset are incorrect. This function is used to match the displayed voltage
with the expected one.
Steps for setting the Load of each channel:
1. Press Utility and choose Output Setup. Press F1 to select CH1Load, or press
F2 to select CH2Load; press it again to select HighZ or *Ω ("*" represents a value).
2. To change the load value, after selecting *Ω, turn the knob to change the value, press
/
direction key to move the cursor left and right; or press the number keys to
input the desired value. Press F3 or F4 to select the unit. The load range is 1Ω
~10KΩ.
Note:
For either of CH1 Output and CH2 Output on the Front panel, the waveform generator
has a fixed 50Ω Series Impendence. No matter what Value the set parameter is, if the real
load is different from the set one, the displayed voltage will not equal the real voltage.

7.2. To Set the Phase Deviation

You can set the phase deviation of the two channels.
1. Press Utility and choose Output Setup, press F3 to select PhaseDev.
2. Press F3 to switch between On/Off.
3. If On is selected, you can set the value of phase deviation. Turn the knob to change
the value, press
number keys to input the desired value in degree, press F4 to select the unit. The
phase deviation range is 0~360°.
/
direction key to move the cursor left and right; or press the
-95-