5. Battery and fuse replacement

Warning! To avoid electrical shock, disconnect all test leads and
any input signals before replacing the battery. Replace only with
same type of battery. A defect fuse is mostly the result of operating
error. Use the following procedure to replacing the battery:
1.
Disconnect all test leads from any live source, turn the rotary
switch to OFF and remove all test leads from the input
terminals.
2.
The battery cover is secured to the rear cover by two screws.
Using a screwdriver, remove the 2 screws from the case and
remove the rear cover.
3.
Remove battery and replace with a new 9 V battery.
4.
Replace the rear cover and reinstall the screws.
5.
While putting in the battery observe the polarity.
6.
Batteries, which are used up dispose duly. Used up batteries
are hazardous and must be given in the, for this being
supposed, collective container.
7.
Important! A defect fuse must be replaced by a fuse with the
same connection value of 250 mA / 300 V or 10 A / 300 V.