Maintenance

Visually inspect the device, hoses, secretion canister and power cord before each use.
Parts which are damaged must be replaced immediately.
The unit does not require any further maintenance.
At least every 12 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS recommends
an inspection according to the manufacturer's specifications.
Reprocessing
Handling of the suction device determines to a large extent its reliability and safety. The hygiene measures described in the
previous chapters are necessary measures for the protection of patients and users and to maintain functional reliability.
Repairs
The following may require repairs from the manufacturer or an authorized service partner. Prior to sending in the device, please
contact your service partner by phone.
Liquids have penetrated the device
Sudden occurrence of unusual noises
Operational and functional disorders which cannot be resolved by means of the hints described in the chapter
"Troubleshooting".

Sending in the device

Remove and properly dispose of consumables.
Clean and disinfect the product and accessories according to the operating instructions.
Enclose used accessories with the device.
Fill in the form QD 434 „Delivery complaint / return shipment" and the respective decontamination certificate.
This form is enclosed with each delivery and can be found at www.atmosmed.com.
The device must be well padded and packed in suitable packaging.
Place the form QD 434 „Delivery complaint / return shipment" and the respective decontamination certificate in an envelope.
Affix the envelope to the outside of the package.
Send the product to ATMOS or to your dealer.
Warranty
ATMOS cannot guarantee an error-free function nor can ATMOS be held liable for damage to people or goods if
non-original ATMOS parts are used,
the information in these operating instructions are disregarded,
assembly, new installations, modifications, extensions and repairs are done by people who are not authorised by ATMOS.