Guidance and manufacturer's declaration – electromagnetic immunity
The [IH18/1] is intended for use in the electromagnetic environment specified below. The customer or the user
of the [IH18/1] should assure that it is used in such an environment
Immunity test
EN 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power sup-
transient/burst
ply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycle
on power supply
40 % U
input lines
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5s
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level
T
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV line(s) to line(s)
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5s
3 A/m
Electromagnetic environment – gui-
dance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humi-
dity should be at least 30 %
Mains power quality should be that
of a typical commercial or hospital
environment.
The electrical fast transient burst
(EFT) is generated by the switching of
inductive loads. Separation between
the equipment and other loads shall
be considered before installation.
Mains filter is required, if necessary.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that of
T
)
a typical commercial or hospital en-
T
vironment. If the user of the [IH18/1]
requires continued operation during
T
)
power mains interruptions, it is re-
T
commended the [IH18/1] be powered
from an uninterruptible power supply
T
)
or a battery.
T
T
)
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
74