Herunterladen Inhalt Inhalt Diese Seite drucken

Beurer IH 18 Gebrauchsanweisung Seite 18

Vorschau ausblenden Andere Handbücher für IH 18:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Operating
approx. 0.8-1.45 bar
pressure
Atomiser fill
min. 2 ml
volume
max. 8 ml
Medicine flow
ca. 0.4 ml/min.
Sound pressure
< 52 dBA
(acc. to DIN EN 13544-1 section 26)
Mains
230 V~; 50 Hz; 230 VA
connection
UK: 240 V~; 50 Hz; 240 VA
Operating
Temperature: +10 °C to +40 °C
conditions
Relative humidity: 10% to 95%
Storage and
Temperature: 0 °C to +60 °C
transport
Relative humidity: 10% to 95%
conditions
Atmospheric pressure: 500 to
1060 hPa
Aerosol
Flow: 5.31 l/min
Properties
Aerosol delivery: 0.171 ml
Aerosol delivery rate: 0.082 ml/min
Particle size (MMAD): 3.16 µm
Subject to technical modifications.
Particle size diagram
Measurements were performed using a sodium
fluoride solution with a "Next Generation Impac-
tor" (NGI).
This diagram may not be applicable for suspensions
or highly viscous medicines. More information can be
obtained from the relevant medicine manufacturer.
Spare parts
Designation
Nebulisation stopper
Yearpack (contains Mouth-
piece, Adult mask, Child
mask, Atomiser, Compressed
air hose, Filter)
Note
If the unit is used outside of the specifications, proper
function is no longer guaranteed!
Material
REF
PP
162.819
PP/ PVC
601.21
18
We reserve the right to make technical changes to im-
prove and further develop the product.
This device and its accessories comply with European
standards EN60601-1 and EN60601-1-2, as well as
EN13544-1, and is subject to special safety measures
in terms of electromagnetic tolerance. Note that port-
able and mobile RF communication equipment can
affect this unit. More details can be requested from
the stated Customer Service address or found at the
end of the instructions for use. The unit conforms to
the requirements of the European Directive for Medical
Products 93/42/EC, the MPG (German Medical units
Act).
ELECTROMAGNETIC COMPATIBILITY
• The device complies with current specifications with
regard to electromagnetic compatibility and is suit-
able for use in all premises, including those des-
ignated for private residential purposes. The radio
frequency emissions of the device are extremely low
and in all probability do not cause any interference
with other devices in the proximity.
• It is recommended that you do not place the device
on top of or close to other devices. Should you notice
any interference with other electrical devices, move
the unit or connect it to a different socket.
• Radio equipment may affect the operation of this
device.