Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance And Care - ergoline Evolution 500 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Evolution 500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Please read and observe ...

Maintenance and care

Extreme danger!
The device must be deenergised when performing maintenance
work requiring the opening of the device.
See Page 45.
Maintenance and care play a major role in deciding whether the device
meets the requirements placed on it in a satisfactory manner. The com-
pliance with the specified maintenance intervals and the careful conduct-
ing of the maintenance and care work are therefore absolutely necessary
– see Page 86 ff.
The current operating hours of the parts requiring maintenance can be
displayed in the Presetting mode – see Page 137 ff.
Remount any protective devices (e.g. filter panels) after completing work.
We point out that the device must be subject to repetitive inspections
by our customer service or an approved specialised company every
12 months (from commissioning) to maintain its proper condition!
Note:
In the Evolution Automatic Power System, the lamp power is
controlled by switching the UV low-pressure lamps. Therefore,
the entire UV low-pressure lamp set must be replaced after the
specified operating time. The new UV low-pressure lamps are
not operable until the related chipcard is inserted (see Page 96).
Caution!
Only use genuine spare parts of the same type!
The use of other parts voids the CE conformity!
Any liability is excluded in case of damage or injury proven to be
the result of the use of non-genuine spare parts.