Fault
is displayed.
StOP
is displayed.
tEMP
The device no longer reacts
when keys are pressed.
10

SERVICING

10.1

Verified scales

checksum
Cause
Maximum capacity has been exceeded.
The ambient temperature of the scale is
too high or too low.
Device is in an undefined state following
implausible input.
The keypad is defective.
CAUTION!
Faulty measurements as a result of verification being performed poorly or not at
all
► Have verification performed only by authorized persons.
► Always have verification performed if one or more quality seals are damaged.
seca recommends having your device serviced prior to verification.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or an au-
thorized service partner.
► You can find a service partner in your vicinity at www.seca.com or by email-
ing service@seca.com.
Have authorized persons perform verification in line with national legal regulations.
Verification is necessary whenever one or more quality seals are damaged or the con-
tents of the verification counter no longer match the number on the applicable verifica-
tion sticker. If quality seals are damaged, contact seca Service directly.
Verifications may only be performed by authorized agencies. To guarantee this, the
scale is equipped with a verification counter which records each change in verification-
related data.
If you want to check whether the scale has been properly verified, proceed as follows:
1. Ensure that the device is switched off.
2. Press and hold any key and start the scale.
ð The current content of the verification counter flashes in the display for a few
seconds.
3. Compare the content of the verification counter output with the number given on the
2
verification counter seal.
Remedy
Remove the load from the device.
• Operate the device only in the intended
ambient conditions (General technical
data).
• After transporting or storing the device,
wait approx. 15 minutes before switching
it on.
• Take power supply unit out of the power
supply socket.
• Wait approx. 1 minute.
• Plug the power supply unit into the
power supply socket; the scale and the
multifunctional display will switch on au-
tomatically.
Inform seca Service.
English
Servicing • 59