10.2
Non-verified scales
11
TECHNICAL DATA
11.1
General technical data
Parameter
Dimensions, total:
• Depth
• Width (incl. ramps)
• Height (operation/transport)
Dimensions, weighing platform:
• Depth
• Width
• Height
Weight (incl. ramps)
Height of digits, display
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
Ambient conditions, storage/transport
• Temperature
• Air pressure
• Humidity
Medical device in accordance with regulation (EU)
2017/745
IEC 60601-1: Medical electrical device, Type B
Type of protection in accordance with IEC 60529
Duty cycle
Service life, from commissioning
Application parts according to IEC 60601-1
60 • seca 665/seca 664
Both numbers have to match for the verification to be valid. If the verification sticker and
the verification counter do not match, the scale must be verified. Please contact your
service partner or seca Service. Once the scale has been verified, a new, updated verifi-
cation counter sticker is used to identify the verification counter reading. The person au-
thorized to perform verification secures this verification with an additional seal. The verifi-
cation counter sticker can be ordered from seca Service.
The product needs to be set up carefully and serviced regularly. Depending on how fre-
quently the product is used, seca recommends servicing at intervals of 3 to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or an au-
thorized service partner.
► You can find your local service partner at www.seca.com.
a
Value
+10 °C to +40 °C / +50 °F to 104 °F
30 % – 80 % no condensation
-10 °C to +65 °C / +14 °F to 149 °F
Class I with measuring function
Weighing platform, display column, display housing, film
1100 mm
1176 mm
912 mm/202 mm
800 mm
965 mm
57 mm
approx. 32 kg
25 mm
700 – 1060 hPa
700 – 1060 hPa
0 % – 95 % no condensation
IP20
Continuous duty
at least 10 years
keypad, display