– Unit:
In the "Unit" menu you can choose whether the tempera-
ture specifications are displayed in "°C" or "°F".
– Audio Signal:
In the "Audio Signals" menu you can switch the audio sig-
nal that sounds upon a surface temperature alarm on and
off.
– Measuring Frame:
In the "Measuring Frame" menu you can switch the meas-
uring frame b on the display on and off.
– Colour Scheme:
In the "Colour Scheme" menu you can choose the colour
of the temperature values and other indications on the dis-
play. The setting will also be applied to saved screenshots.
– Shutdown Time:
In the "Shutdown Time" menu you can choose the time in-
terval after which the measuring tool will automatically shut
down if no button is pressed. You can also deactivate the au-
tomatic shutdown by selecting the "Never" setting. The
shorter the shutdown time, the more energy you can save.
Troubleshooting – Causes and Corrective Measures
Error
Measuring tool cannot be switched on.
Function indicator t for contact tempera-
ture measurement does not appear on the
display.
Measuring tool cannot be connected to a
computer.
Bosch Power Tools
– Light Duration:
In the "Light Duration" menu you can choose the time in-
terval after which the brightness of the display lighting will
reduce if no button is pressed on the measuring tool. You
can also illuminate the display permanently by selecting
the "Always" setting. The shorter the light duration, the
more energy you can save.
– Delete All Images:
In the "Delete All Images" menu you can delete all files in
the internal memory at once. Press the right-hand arrow
button 11 for "more ..." to enter the submenu. Then press
either the left-hand function button 17 under the tick sym-
bol to delete all files, or the right-hand function button 10
under the cross symbol to cancel the operation.
– Tool Information:
Open the "Tool Information" submenu for information
about the measuring tool. There you will find the serial
number of the measuring tool and the installed software
version.
Cause
Corrective Measure
Battery pack or batteries
Charge the battery pack or change the batteries.
empty
Battery too warm or too cold Allow the battery to reach the correct temperature or
change it.
Measuring tool too warm or
Allow the measuring tool to reach the correct temper-
too cold
ature.
Image memory defective
Format the internal memory by deleting all images
(see "Delete All Images", page 29). If the problem
persists, send the measuring tool to an authorised
Bosch after-sales service agent.
Image memory full
If required, transfer the images to another storage
medium (e.g. computer or notebook). Then delete
the images in the internal memory.
Measuring tool is defective. Send the measuring tool to an authorised Bosch
after-sales service agent.
Connection 25 for thermo-
Send the measuring tool to an authorised Bosch
couple defective
after-sales service agent.
Measuring tool not recog-
Check whether the driver on your computer is up to
nised by computer.
date. It may be necessary to have a newer operating
system version on your computer.
Micro USB connection or
Check whether the measuring tool can be connected
micro USB cable defective
to a different computer. If not, send the measuring
tool to an authorised Bosch after-sales service agent.
Humidity and ambient tem-
The other functions of the measuring tool can still be
perature sensor 5 defective
used.
Send the measuring tool to an authorised Bosch
after-sales service agent.