Mode "
1. With the instrument in stand-by, press the
select the desired type of reference (see § 4.2)
2. Press the
point measuring section. The symbol "
on the display with side "1" flashing
3. Position the instrument in the lowest point (1) of
measurement (see Fig. 11 ) and press the
or the side MEAS key. The result appears on the
first partial display. The symbol "
display with side "2" flashing
4. Position the instrument as horizontally as possible
(2) (see Fig. 11) and press the
MEAS key to measure. The result appears on the
second partial display. The symbol "
the display with side "3" flashing
5. Position the instrument in the highest point (3) of
measurement (see Fig. 11) and press the
or the side MEAS key to measure. The result
appears on the second partial display
6. The final value of the result obtained with the
combination of previous measures (see dashed line
in Fig. 11) is shown on the resulting display
7. Measuring result is automatically saved in the
instrument's memory
Fig. 11: Indirect 3-point measurement – Mode 2
"
key three times to enter the indirect 3-
EN - 16
iDM70
key to
" appears
" appears on the
key or the side
" appears on
key