Test of precision

Remove the tube from the manometer and hold the manometer in
a vertical position. When the pointer stands still at 0 on the scale,
the instrument has been adjusted properly. If the pointer is below
or above 0, the instrument must be recalibrated. Either take it to
an authorised RIESTER dealer or send it to us.

Monitoring of instruments

EU countries except Germany:
Legal regulations for monitoring instruments apply to all EU coun-
tries except Germany.
Countries outside the EU:
For any countries where no legal regulations exist for monitoring
instruments, it is recommended to examine the accuracy of mea-
suring instruments in 2-year intervals.

Technical Data

Minimal environmental conditions under which the error tolerance
of +/- 3 mm Hg must be satisfied in accordance with the EN-
1060-2 standard:
Measurement conditions: 10°C to 40°C at a relative air
Storage conditions:
Versions:
Type of indication:
Scale graduation:
Range shown on scale:
Measurement range:
Pointer movement:
Tube adapter:
Pressure generation:
Pressure reduction:
humidity of 85 % (non-condensing)
-20°C to 70°C at a relative air
humidity of 85 % (non-condensing)
Aneroid, desk, wall, stand and
anaesthetic model
Round scale
Increments of 2 mm Hg
0 to 300 mm Hg
0 to 300 mm Hg
No stop pin
"1 or 2" or "1 and 2",
depending on the particular model
Bulb
Air-release valve that can be regu-
lated.
Attention, please see instruction manual
Key to symbol on the unit:
Meaning of the symbol on the packaging: