11. Technical specifications
Operating
10 - 40 °C / 50 - 104 °F
conditions:
15 - 90% relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 90% relative maximum humidity
Weight:
124 g (including batteries)
Dimensions:
131 x 90 x 60.5 mm
Cuff size:
13.5 - 21.5 cm (5.25 - 8.5 inches)
Measuring
oscillometric, corresponding to Korotkoff
procedure:
method: Phase I systolic, Phase V diastolic
Measurement range: SYS: 60 - 255 mmHg
DIA: 40 - 200 mmHg
Pulse: 40 - 199 beats per minute
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
within ± 3 mmHg
Pulse accuracy:
± 5% of the readout value
Voltage source:
2 x 1.5 V alkaline batteries; size AAA
Microlife BP W1 Basic
Battery lifetime:
approx. 320 measurements
(using new batteries)
IP Class:
IP 22
Reference to
IEC 80601-2-30; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000
measurements, whichever comes first
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.