This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage
caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical specifications» section.
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device if you think it is damaged or notice anything
unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
The measurement results given by this device is not a diagnosis.
It is not replacing the need for the consultation of a physician,
especially if not matching the patient's symptoms. Do not rely on
the measurement result only, always consider other potentially
occurring symptoms and the patient's feedback. Calling a doctor
or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Microlife BP W1 Basic
Contra-indications
Do not use this device if the patient's condition meets the following
contra-indications, to avoid inaccurate measurements or injuries.
The device is not intended for measuring blood pressure in pedi-
atric patients of age younger than 12 years old (children, infant,
or neonates).
Presence of significant cardiac arrhythmia during measurement
may interfere with blood pressure measurement and affect the
reliability of blood pressure readings. Consult with your doctor
about whether the device is suitable for use in this case.
The device measures blood pressure using a pressured cuff. If
the measuring limb suffers from injuries (for example open
wounds) or under conditions or treatments (for example intrave-
nous drip) making it unsuitable for surface contact or pressuriza-
tion, do not use the device, to avoid worsening of the injuries or
conditions.
Patient motions during measurement may interfere with the
measurement process and influence results.
Avoid taking measurements of patients with conditions, diseases,
and susceptible to environment conditions that lead to incontrol-
lable motions (e.g. trembling or shivering) and inability to commu-
nicate clearly (for example children and unconscious patients).