Log in to SIGRIST-Webinterface
Log in without password.
u
Recommendation: Secure access to the
SIGRIST-Webinterface with a password.
7
Troubleshooting
7.1
Isolate faults
Malfunction
No display
Error message in display
Measured value seems
wrong
7.2
Warning/(Prio) error messages
In the event of a malfunction, either the measurement monitor with error
sponding status
With a swipe movement, the detailed information appears.
Warning messages
System remains in operation.
l
Evaluate measurement results with cau-
l
tion.
The warning disappears after the cause
l
has been rectified.
Call up QR code (5).
u
Rectify cause promptly.
u
(1)
Warning message with measured value
display
(2)
Warning status symbol
(3)
Warning code
(4)
Warning message
(5)
QR code
32 / 208
Troubleshooting
Measure
Check service voltage.
u
Analyse error message (Warning/error/priority messages).
u
Ensure correct operating conditions of the sample medium.
u
Check calibration.
u
Check correct mounting.
u
Ensure that servicing duty has been carried out correctly.
u
Perform sensor check.
u
signal (2)
is displayed, depending on the setting.
1
3
4
5
message (1)
or a corre-
2