The device must not be operated directly beside or on top of other devices. If the medical
electrical device must be operated beside or on top of other devices, then it must be mon-
itored to ensure that it is operating properly in the arrangement used.
On request, technical data on electromagnetic compatibility (EMC information) is available in
table format from the manufacturer or distributor, or on the website [see: Links, page 39]
Ambient conditions
DURING OPERATION
Ambient temperature
Relative humidity
Atmospheric pressure
The compressor is intended for operation in all healthcare situations. It is not permitted to
operate it in trains, motor vehicles or aeroplanes.
Use of the compressor in professional healthcare facilities is limited to the inpatient wards
and the intensive care unit. Use of the compressor in areas with elevated magnetic or elec-
trical radiation (e.g. close to an MRI scanner) is not permitted.
TRANSPORT AND STORAGE BETWEEN USES
Minimum ambient temperature (without monitoring of relative
humidity)
Maximum ambient temperature (with relative humidity of up
to 93%, non-condensing)
Humidity
Atmospheric pressure
6.2 Nebuliser
General nebuliser data
8
Size
8
Weight
Operating gases
Minimum compressor flow
Minimum operating pressure
Maximum compressor flow
8) Without mouthpiece and mask; unfilled.
+10 °C to +40 °C
30% to 75% (non-condensing)
700 hPa to 1,060 hPa
-25 °C
+70 °C
max. 93%
500 hPA to 1,060 hPA
10 cm × 9.5 cm × 4.5 cm
27.5 g to 29.5 g
Air, oxygen
3.0 l/min.
0.5 bar / 50 kPa
6.0 l/min.
37