Elastic band
Compressor housing
Air filter
BEFORE A CHANGE OF PATIENTS
Nebuliser components
Mask components
Connection tubing
Elastic band
Compressor housing
Air filter
Reprocessing for patients at risk
CAUTION
Danger of infection for patients at risk
For patients at risk, airway infections represent an increased risk of deterioration of gen-
eral health, because these patients are particularly vulnerable to residual germs. Patients
at risk include cystic fibrosis patients, patients with immunosuppression or immunodefi-
ciency, and vulnerable patient groups.
• If you are a patient at risk, disinfect the individual parts once per day during the therapy.
• If you are unsure whether you are a patient at risk, consult with specialist medical per-
sonnel before use.
Cleaning visible dirt
Cleaning visible dirt
Replace after 200 operating hours (approx. 1 year)
– Cleaning
– Disinfection
– Sterilisation
– Cleaning
– Disinfection
– Sterilisation
Mechanical cleaning with disinfection
Replacing the elastic band
– Cleaning
– Wipe disinfection
Replace after 200 operating hours (approx. 1 year)
6
6
27