4.6 Reprocessing the connection tubing
Dry the connection tubing after each use.
The maximum operating life of the connection tubing is 1 year.
Procedure
• Connect the connection tubing to a compressed air source (compressor or cent-
ral medical gas supply system).
• Switch the compressed air source on.
• Leave the compressed air source running until all the moisture in the tube has
been removed.
• Inspect all product components after each cleaning and disinfection.
• Replace defective, deformed or seriously discoloured parts.
• Store the product as described below:
– Dry and dust-free, e.g. in a clean, lint-free cloth (e.g. a tea towel)
– Protected against contamination, where necessary (e.g. using optional sterile
packaging)
4.7 Further information about reprocessing
Further validated processes for reprocessing
The instructions provided were validated by PARI and were found to be suitable for prepar-
ing your medical device for its reuse.
34
EQUIPMENT:
The method has been validated in Europe using:
®
– Neodisher
MediClean forte (concentration: 0.5%)
– Deionised water
– Cleaning and disinfection device as per ISO 15883.
– Special baskets for Miele instrument dishwasher
– Compressed air source for drying
PROCEDURE:
• Programme for cleaning and thermal disinfection (at least A0 = 3000)
as per manufacturer's instructions.