4.2 Limits of reprocessing

Limits of reprocessing in a home environment
Nebuliser and accessories, disinfection
Mask, disinfection
Limits of reprocessing in a professional healthcare institution
Nebuliser and accessories, disinfection
Nebuliser and accessories, sterilisation
Connection tubing
Mask, disinfection
Mask, sterilisation

4.3 Preparation for reprocessing

Procedure
• Detach the connection tubing from the compressor.
• Detach the connection tubing from the nebuliser.
• Detach the mouthpiece from the nebuliser.
• Make sure that all residual volume is removed from the nebuliser.
• Dismantle the nebuliser into its individual parts.
• Carefully pull the blue exhalation valve out of the slot in
the mouthpiece. The exhalation valve must still be at-
tached to the mouthpiece.
• Disconnect all mask components from the nebuliser.
• Dismantle the mask into all its individual parts.
28
300 processing cycles, max. 1 year
300 processing cycles, max. 1 year
300 reprocessing cycles, max. 1 year
100 processing cycles, max. 1 year
50 reprocessing cycles,
max. 1 year
300 processing cycles, max. 1 year
100 processing cycles, max. 1 year