4.4 Reprocessing of nebuliser and mask
The following products can be cleaned, disinfected and sterilised according to the procedure
described below:
– PARI nebuliser and PARI accessories
– PARI mask soft
The connection tubing and elastic band of the mask must be treated separately.
Procedure
All individual parts must be precleaned immediately after use.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
EQUIPMENT:
– Drinking water temperature of about 40 °C
– Standard commercial washing-up liquid
– Container having sufficient capacity
PROCEDURE:
• Unless otherwise specified by the manufacturer
of the washing-up liquid, add about 1 teaspoon
of washing-up liquid to 3 l warm drinking water.
• Place all the parts in the washing-up water.
Application time: 5 minutes
• Occasionally move the parts back and forth.
• In case of visible soiling, use a medium-soft brush (e.g. a toothbrush)
which is used exclusively for this purpose.
RINSING:
• Rinse off all parts thoroughly in running drinking water at about 15 °C
for 3 minutes.
Cleaning the elastic band
• Clean the elastic band as necessary with warm drinking water and a
little dishwashing liquid.
The elastic band cannot be disinfected or sterilised.