Herunterladen Diese Seite drucken

Pari COMPACT2 Gebrauchsanweisung Seite 82

Inhalationssystem
Vorschau ausblenden Andere Handbücher für COMPACT2:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Maximum operating pressure
Minimum fill volume
Maximum fill volume
Aerosol characteristics according to ISO 27427
The aerosol characteristics presented in these instructions for use were determined in ac-
cordance with ISO 27427 using 2 ml Salbutamol fill volume. If other solutions or suspensions
are used for nebulisation, the aerosol characteristics may differ from the values shown (par-
ticularly if they have greater viscosity).
The following data is based on tests according to a standard which takes adult breathing pat-
terns as a basis. Therefore, these figures will probably differ from corresponding figures that
were calculated for populations of children and infants.
Nozzle insert
(transparent)
10
MMAD [µm]
11
GSD
Respirable fraction
[% < 5 µm]
Aerosol fraction [% < 2 µm] 11.6
Aerosol fraction [% > 2 µm
< 5 µm]
Aerosol fraction [% > 5 µm] 56.0
Aerosol output [ml]
Aerosol output rate [ml/
min]
Residual volume [ml] (gra-
vimetric)
Percentage of fill volume
emitted per minute [%/min]
9) Operation with PARI COMPACT2 compressor (Type 152).
10) MMAD = Mass Median Aerodynamic Diameter
11) GSD = Geometric Standard Deviation
38
Minimum
compressor flow
(3 l/min – 0.6 bar)
5.4
2.13
44.0
32.4
0.30
0.09
1.28
4.4
2.0 bar / 200 kPa
2 ml
8 ml
Nominal
compressor flow
(4 l/min –
9
1.2 bar)
4.5
2.23
53.2
17.9
35.3
46.8
0.40
0.12
1.15
5.9
Maximum
compressor flow
(6 l/min – 1.9 bar)
3.4
2.31
66.2
28.1
38.1
33.8
0.38
0.14
1.14
6.9

Werbung

loading

Diese Anleitung auch für:

Compact2 152Lc plus 022041