NOTE
Risk of damage to plastic parts!
Plastic melts if it comes into contact with hot surfaces.
• During thermal disinfection, you must ensure that the water level in the container is suf-
ficient, in order that the individual parts do not come into contact with hot surfaces.
Always use drinking water for reprocessing in a home environment.
In professional health care institutions, drinking water is sufficient for precleaning. For
all other reprocessing steps and processes, use deionised water with a low microbiolo-
gical load (at least drinking water quality).
Please observe the instructions for use for the chemicals used.
Wash your hands thoroughly before every reprocessing.
4.1 Reprocessing cycles
Perform the reprocessing steps as per the specified reprocessing cycles.
Reprocessing cycles in a home environment
Nebuliser components
Mask components
Elastic band
Compressor housing
Air filter
Reprocessing cycles in professional healthcare institution
SINGLE PATIENT USE
Nebuliser components
Mask components
5) For the therapy of patients at risk, disinfect the individual parts once per day. Take note of the
additional information for patients at risk [see: Reprocessing for patients at risk, page 27].
6) Air filters are included in every PARI Year Pack.
26
– Clean immediately after every use
– Disinfect once per week
– Clean immediately after every use
– Disinfect once per week
Cleaning visible dirt
– Cleaning in the event of visible soiling and before each
patient change
– Disinfect by wiping before each patient change
Replace after 200 operating hours (approx. 1 year)
– Clean immediately after every use
– Disinfect once per week
– Clean immediately after every use
– Disinfect once per week
5
5
5
5
6