Procedure
• Inspect all product components after each cleaning and disinfection.
• Replace defective, deformed or seriously discoloured parts.
The air filter must be checked at regular intervals (after every 10th use). If it is dis-
coloured brown or grey, or if it is damp or clogged, it must be replaced.
The air filter cannot be cleaned and then reused!
Removing the air filter
• Pull the filter holder out of the compressor. Use a small
screwdriver, for example, to prise the filter holder care-
fully out of the compressor.
Replacing the air filter
• NOTE! For operating your compressor, use only air filters provided by the man-
ufacturer or distributor. If air filters not designed for the compressor are used, it
may be damaged.
Pull the old air filter out of the filter holder, and fit the new filter in its place.
• Insert the filter holder back in the compressor.
•
CAUTION! Electrical devices connected to the
power supply represent a potential source of danger.
When storing the compressor, always pull the power
plug out of the socket.
• NOTE! Do not wind the power cord around the compressor. If the power cord is
wound or bent very tightly, the wires inside the cord may break. The power cord
will then be unusable.
Wind the power cord up loosely.
• Secure the cable end in the cable holder.
• Wrap all individual components in a clean, lint-free cloth (e.g. a tea towel).
• Store the product in a dry, dust-free place.
33