Cleaning & Disinfection
The device may be wiped down with a standard disinfectant wipe.
Never spray disinfectant directly onto the device.
CAUTION: Always use
a protective barrier.
Black nosecone is removable and autoclavable.
Between uses, check the lens to ensure it is clean.
A dirty lens will decrease performance.
The Phasor™ is provided non-sterile. There are no special accessories needed
to sterilize the Phasor™ heated composite delivery system.
Quaternary ammonium compound products are recommended
(containing 20% alcohol or less). Wipe, do not spray, solution onto the unit.
Prevent liquids from entering openings on the composite gun unit.
DO NOT AUTOCLAVE HANDPIECE.
REMOVABLE NOSECONES CAN WITHSTAND
AUTOCLAVE - (132°C / 0.22mPA for 3 minutes).
8
CAUTION:
• DO NOT immerse the unit or unit parts in solutions. Use of
solutions other than those recommended may damage plastic
parts and will void product warranty.
• DO NOT use abrasive material such as scouring powder, organic
solvents, or solvent-based cleaning fluids. In case of severe
contamination, gently clean the device by using diluted alcohol.
• Store the device in the box if it is not to be used for an extended
period of time.
Infection Control Measures
CAUTION: To prevent cross-contamination, a disposable plastic
sleeve must be used over the Phasor™ with each use. A low-density
polyethylene plastic disposable barrier covers the nose cone of unit
and provides a hermetically sealed barrier between the handpiece
and patients. The disposable barrier limits patient-to-patient
contamination. Discard used barrier sleeves after each patient.