Herunterladen Diese Seite drucken

Hettich ROTOFIX 32 A Gebrauchsanweisung Seite 70

Vorschau ausblenden Andere Handbücher für ROTOFIX 32 A:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Software operation
7.1.4
Centrifuging radius
7.2
Rotor detection
7.3
Machine Menu
7.3.1
Audible signal
7.3.1.1
General
28 / 38
For example: Maximum speed 4000 RPM, density 1.6 kg/dm³
If, in exceptional cases, the maximum load indicated on the bucket is
exceeded, the speed must also be reduced. The permissible speed can
be calculated using the following formula:
For example: Maximum speed 4000 RPM, maximum load 300 g, actual
load 350 g
Please contact the manufacturer if you are not sure.
The centrifuging radius must be entered in centimetres.
1.
Switch off the mains switch.
2.
Press and hold down the ▲
[IMPULSE] button simultaneously.
3.
Switch on the mains switch and release the buttons.
[RPM/RCF x 100] button repeatedly until the centrifuging
4.
Press the ▲
'rd' are displayed.
radius and
5.
Use the
[t] buttons to set the desired centrifuging radius.
[STOP] button to save the settings.
6.
Press the
Rotor detection is performed after starting a centrifugation run.
If the rotor has been changed, the centrifugation run is cancelled after
rotor detection. The rotor code (red) is displayed.
If the maximum speed of the rotor used is less than the set speed, the
speed is limited to the maximum rotor speed.
The audible signal sounds:
after a problem occurs in the 2 s interval.
after completion of the centrifugation run and rotor standstill in the 30 s
interval.
Opening the lid or pressing any button stops the audible signal.
Rev.: 00 / 11.2023
[RPM/RCF x 100] button and the
AB1207en

Werbung

loading