4
Transport and storage
4.1
Transport and storage conditions
Transport conditions
Storage conditions
4.2
Fastening the transport lock
AB1207en
The original transport locks must be attached for return shipment, see
Chapter 4 'Transport and storage' on page 15 . The device must be
⮫ 
shipped in its original packaging.
NOTICE
Damage to the device due to failure to use the transport
locks.
Secure the transport locks before transporting the
device.
NOTICE
Damage to the device due to condensation.
There is a risk of condensation forming on electrical com-
ponents when component surfaces are cold and the sur-
rounding air is warmer. The condensation that forms may
cause a short circuit and/or destroy electronics.
Warm the device up for at least 3 hours in a warm room
before connecting it to the mains.
or
Warm up for 30 minutes in a cold room.
Before transporting, fasten the transport lock and disconnect the device
from the mains socket.
The transport temperature must be between -20 °C and +60 °C.
Humidity must not be condensing. Humidity must be between 10% and
80%. 
Be aware of the weight of the device.
When transporting using a transport aid (e.g., a pallet truck), the trans-
port aid must be able to carry at least 1.6 times the transport weight of
the device.
Secure the device to prevent it tipping over and falling down during
transport.
Never transport the device sideways or upside down.
The device must be stored in the original packaging.
Only store the device in dry rooms.
The storage temperature must be between -20 °C and +60 °C.
Humidity must not be condensing. Humidity must be between 10% and
80%. 
Personnel:
Trained user
The lid is closed.
The mains cable is disconnected from the device.
Rev.: 00 / 11.2023
Transport and storage
15 / 38