These documents can be found:
S On the QuantumX system CD supplied with the device
S After installing the QuantumX Assistant on the hard
drive of your PC
S Up-to-date versions are always available from our In
ternet site at
Note
When working with the MX403B, always take note of the
standard operating manual (I2322) as well.
The high level of protection of the MX403B is guaranteed
by consistent development in accordance with the latest
conditions specified in the measuring equipment stand
ards IEC 61010-1:2010+Cor.:2011 and
IEC 61010-2-030:2010+Cor.:2011, as well as by VDE
certification and production monitoring.
The MX403B has four insulated differential measurement
channels for directly measuring voltages up to 1000 V DC
or 1000 V rms AC.
The different measuring ranges allow the acquisition of
high voltages to remote ground, as well as the measure
ment of small differential voltages at high potential to re
mote ground.
The connection is made with 4 mm safety lab jacks,
19 mm pitch, suitable for standardized BNC adapters.
Each channel is equipped with a programmable amplifier,
analog reconstruction and antialiasing filters, a 24bit
ADC and digital filters.