S The measurement module and devices used in auto­
mation must be covered over in such a way that ad­
equate protection or locking against unintentional ac­
tuation is provided (e.g. access checks, password
protection, etc.).
S For those measurement modules operating in net­
works, safety precautions must be taken in terms of
both hardware and software, so that a line break or
other interruptions to signal transmission do not cause
undefined states or loss of data in the automation
device.
S After making settings and carrying out activities that
are password-protected, ensure that any controls that
may be connected remain in a safe condition until the
switching performance of the measurement module
has been tested.
S The maximum cable length must not exceed 30 m.
It is not acceptable to run cables through outdoor
areas.
S Measurement signals can only be applied to the
MX403B once it has been connected to a QuantumX
application (such as the QuantumX Assistant).
DANGER
Voltages with no energy limit that exceed one of the fol­
lowing values are rated as dangerous according to EN
61010:
1. AC voltage, 33 V rms value
2. AC voltage, 46 V peak value
3. DC voltage, 70 V
MX403B
A3757-1.0_en/de/fr