Electromagnetic compatibility
Table 206:
Recommended separation distances between portable and mobile RF communications equipment and
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The operator of
the massaging unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
the equation applicable to the frequency of the transmitter, where
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHzl
,
d
=
1 2 P
0.12
0.38
1.2
3.8
12
is the maximum output power rating of the transmitter in watts (W) according to
P
m
80 MHz to 800 MHz
,
d
=
1 2 P
0.12
0.38
1.2
3.8
12
in metres (m) can be estimated using
d
800 MHz to 2.5 GHz
,
d
=
2 3 P
0.23
0.73
2.3
7.3
23