Electromagnetic compatibility

Table 201:
Guidance and manufacturer's declaration – electromagnetic emissions
is intended for use in the electromagnetic environment specified below. The operator of this unit should assure that it is
used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-11
46
Compliance
Electromagnetic environment – guidance
for its internal function. Therefore, its RF emis-
Group 1
sions are very low and are not likely to cause
any interference in nearby electronic equip-
ment.
Class B
all establishments, including domestic estab-
lishments and those directly connected to the
public low-voltage power supply network that
Class A
supplies buildings used for domestic purposes.
Obligations according to Subclause 6.2.2 of
EN 61000-3-11:2000 have to be considered.
Complies
Especially the line impedance of 0.14 Ohms
must be guaranteed for pulse massaging.
uses RF energy only
is suitable for use in