Technical data
9
Technical data
9.1
General data
Parameter
Design
Measuring principle
Measuring range
Precision within the measuring range
0 ... 1 mS/cm
0 ... 10 mS/cm
0 ... 50 mS/cm
0 ... 100 mS/cm
0 ... 1000 mS/cm
0 ... 2000 mS/cm
Cell constant
Temperature sensor
t
Temperature
a
90
Permissible ambient temperature
Permissible storage temperature
Degree of protection
Permitted medium temperature
when operating in pipework made
of:
Maximum process pressure
at 20 °C
at 60 °C
at 80 °C
30
For operation in the flow
b
ICT 5
inductive
0 ... 2000 mS/cm
k = 6.25 1/cm
Pt1000, Class A
≤ 6 seconds
-10 ... +60 °C
-20 ... +75 °C
PVC: -10 ... +60 °C
PP: -10 ... +80 °C
with PP design
10 bar
6 bar
0 bar
ICT 5-IMA
For immersion
≤ 1%
≤ 0.5%
≤ 0.5%
≤ 0.5%
≤ 1%
≤ 1%
IP68
-10 ... +60 °C
0 bar
0 bar